Aquasearcher™ AB33EC Bench Meter, Ohaus
Conductivity meter, AB33EC-F, AB33EC-B content; STCON7 probe; conductivity standard kits (84 µS/cm, 4×50 ml)
Catalog # 470336-590
Jump to:
Product Details & Documents
Highly reliable conductivity/TDS/salinity/resistivity benchtop meter for standard laboratory applications.
- Selectable reading reference temperatures of 20 or 25 °C with linear curve options and automatic temperature compensation ensures accurate results
- Features such as auto-stop, auto temperature compensation, adjustable TDS factor, 2- or 4-cell conductivity probe compatibility make AB33EC well suited for universal laboratory applications
- A 1000-item memory for conductivity measurements and calibration trail allows for efficient data documentation
- Standard RS232 and USB interface allows connection to external devices
The next step in the evolution of OHAUS’ original Starter Series, the all-new AQUASEARCHER™ AB33EC bench meter is designed with reliable measurement and user-interface in mind. The i-Steward ensures repeatable and consistent measurement accuracy for optimal peace of mind. The 6.5 inch LCD display and touch keypad makes changing the parameters, executing setup, and performing calibration as simple as using a smartphone.
Certifications: Product Safety: IEC/EN 61010-1. Electromagnetic Safety: IEC/EN 61326-1 Class B, basic environments; FCC Part 15 Class A; Canada ICES-003 Class A. Compliance Marks: CE, RCM. Others: WEEE, RoHS.
Specifications
- Description:Conductivity meter, AB33EC-F
- Includes:AB33EC-B content; STCON7 probe; conductivity standard kits (84 µS/cm, 4×50 ml)
- Connections:RS232, USB Device (included)
- Cat. No.:470336-590
- Conductivity Range:0.01...19.99 μS/cm;
20...199.9 μS/cm;
200...1999 μS/cm;
2.00...19.99 mS/cm;
20.0...199.9 mS/cm - Conductivity Resolution:0.001 μS/cm min; auto-range
- Parameter:Conductivity
- Conductivity Accuracy:±0.5% reading ± 2 LSD (least signficant digit)
- Display:LCD with backlight and one-line dot matrix